Section 85 in The Legal Metrology (General) Rules, 2011
85.
% at upper limit temperature.Temperature-humidity sequence : Reference temperature of 200°C at 50% humidity;The upper limit temperature at 85% humidity;Reference temperature of 20°C at 50% humidity.Number of test Cycles: At least one cycle.Weighing test and test sequence: After stabilization of the EUT at reference temperature and 50% humidity, apply at least five different test loads or simulated loads and record:(a)date and time;(b)temperature;(c)relative humidity;(d)test load;(e)indications (as applicable);(f)errors;(g)functional performanceIncrease the temperature in the chamber to the upper limit and increase the relative humidity to 85%. Maintain the EUT at no load for a period of 2 days (48 hours). Following the 2 days, apply at least five test loads and record the data as indicated above. Allow full recovery of the EUT before any other tests are performed.Maximum allowable variations: All errors shall be within the maximum permissible errors specified in Table II.| Environmental phenomena | Test specification |
| Voltage variation | Reference voltage |
| Reference voltage + 10% | |
| Reference voltage - 15% | |
| Reference voltage |